Feature | Specification |
---|---|
Measuring scales | °Optical Rotation, °Specific Rotation, %Concentration, user-defined scales |
Measuring range | ±89.9° |
Resolution | 0.001° |
Accuracy | ±0.004° |
Repeatability | ±0.004° |
Response time | max. 12 sec. |
Wavelength | 589 nm |
Light source | LED, with average lifetime of 100 000 hours |
Sensitivity | Light intensity control compensates attenuation up to Optical Density (OD) 2.0 |
Temperature sensor | Built-in PT 100 sensor |
Temperature resolution | 0.1 °C |
Temperature accuracy | ±0.1 °C |
Temperature control | Automatic Peltier temperature control between 15 °C and 35 °C, or without temp. control |
Sample cell / quartz control plate | Toolmaster™ automatic sample cell and quartz control plate identification; sample cell path length of 2.5 mm to 100 mm |
Display | TFT touchscreen 5.7″ (145 mm), 640 × 480 Pixels |
Interfaces | 3 USB, RS232, Ethernet, CAN bus, 1 USB type “B” for service Easy connection of keyboard, mouse, printer, bar code Network capability |
Dimensions (L×W×H) | 370 mm × 320 mm × 130 mm (14.6 in × 12.6 in × 5.1 in) |
Weight | 8.6 kg (19 lbs) |
Power supply / consumption | AC 100 to 240 VAC, 50/60Hz / 70VA, 120 VA max. |