| Model | SVM 2001 |
| Patents | AT516058 (B1), US10036695 (B2), CN105424556, EP2995928 (B1) |
| Test methods | ASTM D7042, D445 bias-corrected, ISO 23581, EN 16896, ASTM D4052 (optional) |
| Measuring Range | |
| Viscosity | 0.2 mm²/s to 30,000 mm²/s* |
| Density | 0 g/cm³ to 3 g/cm³ |
| Temperature | +15 °C to +100 °C |
| Precision | |
| Viscosity repeatability | 0.1 % |
| Viscosity reproducibility | 0.35 % |
| Density repeatability | 0.0002 g/cm³ (0.00005 g/cm³ with ASTM D4052) |
| Density reproducibility | 0.0005 g/cm³ (0.0001 g/cm³ with ASTM D4052) |
| Temperature repeatability | 0.005 °C |
| Temperature reproducibility | 0.03 °C from +15 °C to +100 °C |
| Performance | |
| Sample volume min./typical | 1.5 mL / 5 mL |
| Solvent volume min./typical | 1.5 mL / 6 mL |
| Max. sample throughput | 33 samples per hour |
| Features | |
| Optional upgrades | Automatic VI determination, ASTM D4052 density measurement |
| Accessories | Magnetic particle trap, hot filling attachment |
| Automation | Optional sample changer + HTV platform |
| Technical Data | |
| Data memory | 1000 results |
| HID | 10.4″ touchscreen, optional keyboard, mouse, 2D barcode reader |
| Interfaces | 4x USB, 1x Ethernet, 1x CAN, 1x RS-232, 1x VGA |
| Power supply | AC 100–240 V, 50/60 Hz, 250 VA max. |
| Ambient conditions | 15–35 °C, max. 80% RH non-condensing |
| Net weight | 15.9 kg |
| Dimensions (W x D x H) | 33 x 51 x 23.1 cm |
| Trademarks | SVM, FillingCheck, Abbemat |